Matters Needing Attention In The Use Of Rich Packing Laboratory Tablet Press

Mar 14, 2022
When the small laboratory tablet press is not used for a long time, the oil drain valve should be slightly tightened before use. When the pressure increases to 5-10MPa, the load will be unloaded. Repeat 2-3 times in a row for normal use;

The large piston 19 should not exceed the specified stroke, otherwise the tension spring will be deformed and the cylinder cannot return to the initial position.

The tablet lever 20 cannot be shaken, the pressure gauge is not pressed, the screw 26 is loosened, and the low pressure valve port is manually blocked. If the lever 20 is shaken, the oil will flow out from the 26 position and the hand will not block the valve oil. When it comes out, screw back screw 26 and tighten until dead


Automatic Medicine Tablet Press Machine


In order to prevent the tablet press from leaking oil during transportation, the oil injection hole screw was tightened, but the oil collecting tank was well sealed, so the oil cylinder could not inhale hydraulic oil, nor could it pressurize upward normally. During normal use, you need to loosen the screws slightly;

When pressurizing, be aware that the threads must go into the upper plate, not just attach a few buckles. Prevent thread deformation; add lubricating oil to screw 2 regularly. Prevent the screw from rusting and affect its use; the pressure should not exceed the pressure range of the machine, otherwise it will be dangerous; the small laboratory tablet press should use clean No. 46 anti-wear hydraulic oil. When pressure is felt, the manual pressure handle 20 is firm, but the pressure gauge 21 has no indication, and the pressure gauge 21 should be removed immediately.